| HARSHAW
TLD AUTOMATIC SYSTEM

BASIC CAPABILITIES
Dosimeter types: * chips 1/8”
x 1/8” Face thickness from 0.01” to 0.035”
* Rods 1mm dia x 6mm long and 1x1x6 mm
* Disks 1/8” diameter, 0.015” to 0.035 thick
* Cubes 1mm x 1mm x 1mm
Dosimeter Material: * TLD 100,
TLD 200, TLD 300, TLD 400, TLD 600,
TLD 700, TLD 800
Capacity: * 50 Dosimeters in
one loading
Throughput: * Cycle time per chip; 30 seconds with
normal time-temp
profile adjustable by menu from 20-300 seconds.
Heating Method: * Hot gas –
Linear with temperature reproductability better
than +/- 10C
Heating Profile:
a) Adjustable by menu. Up to
ten heating profiles stored and
accessible for viewing.
b) Pre-read anneal; adjustable
0 to 100 second temperature
to 200 0C
c) Linear Ramp; adjustable
from 10C per second to 50oC per
second
d) Maximum Temperature 40oC
e) Post read anneal; 0 to
20 second; up to 400oC
Calibration unit: * User selectable
therefore uGy, cGy, mRem, mR, uSv
APPLICATIONS
a) Radiotherapy planning and evaluations
b) Critical organ dose tracking during radiotherapy
c) Diagnostic studies
d) Medical Research
e) Dose measurements during whole body exposures
f) CT dose measurements for QA
g) Personal Radiation Protection
h) Environmental Dosimetry
OPERATING
MODES
a) Read Dosimeters
b) Calibrate Dosimeters
c) Calibrate Reader
d) QA procedures
e) Reader Setup
|